Atomic Force Microscopy (AFM)
Atomic force microscopy (AFM) or scanning force microscopy (SFM) is a very high-resolution type of scanning probe microscopy (SPM), with demonstrated resolution on the order of fractions of a nanometer, more than 1000 times better than the optical diffraction limit. Using an atomic force microscope (AFM), it is possible to measure a roughness of a sample surface at a high resolution, to distinguish a sample based on its mechanical properties (for example, hardness and roughness). Vast of aims can be obtained with AFM's multiple working mode:
- Scan modes
- Contact mode
- Non-contact mode
- Intermittent-Contact mode
- Tapping mode
Machine Configuration
- N9425A Series 4500 AFM/SPM
- N9512A AFM scanner, 30um, 630nm-Series 4500
- N9740A AFM/LFM detector and cantilever holder -Series 4500
- N9601A AFM/SPM Controller -Series 4500 or 5100
- N9600-80003 LCD Monitor flat-panel 19 inch
- 8120-1395 Agilent qualified, OPT-903 18-AWG 3-COND 96-IN-LG
- OEM Software
- PicoScan Controller 2100 Series
- Computer
AFM Step by Step Instruction
- Load Tip on the tip mount, make sure one thirds of the trip inside the copper clip.
- Amount the scanner and fasten two back screws (finger tight).
- Plug the power cable on the left slot on the hood
- Put the opaque glass block into the detector slot, adjust the laser, make sure it falls on one of the four cantilevers.
- Take out the glass block and insert the detector.Make the signal around -2.5, and the amplitude at least 100.
- Carefully load the sample, and manually approach the sample plat as close as possible.
- Use the software: Picoscan
Set up:
Servo Gain (I&P):0.5
Servo range: Max
Force set point: 0
Motor speed: approaching 10nm
withdrawing 50nm
Scan range for this scanner: 37um*37um
Scan resolution: 512*512